A method for coincidence timing resolution enhancement


Ermis E. E., Celiktas C., Pilicer E.

REVIEW OF SCIENTIFIC INSTRUMENTS, vol.87, no.5, 2016 (SCI-Expanded) identifier identifier identifier

  • Publication Type: Article / Article
  • Volume: 87 Issue: 5
  • Publication Date: 2016
  • Doi Number: 10.1063/1.4948925
  • Journal Name: REVIEW OF SCIENTIFIC INSTRUMENTS
  • Journal Indexes: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Bursa Uludag University Affiliated: Yes

Abstract

A method including the coincidence time resolution improvement for a TOF/positron emission tomography system was suggested. The spectrometer for this aim was composed of two NaI(Tl) and two plastic scintillation detectors. Experimental results were supported by FLUKA Monte Carlo simulation program by constructing the detector setup in software medium. Present experimental results verified our previous results and conclusions obtained from the suggested method. It was concluded that better resolutions would help the improvement not only on the TOF gain but also on the spatial resolution, leading to better images and helping the Physician in his/her diagnosis and treatment. Published by AIP Publishing.