New experimental validation of the pulse height weighting technique for capture cross-section measurements
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, cilt.521, sa.2-3, ss.454-467, 2004 (SCI-Expanded, Scopus)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 521 Sayı: 2-3
- Basım Tarihi: 2004
- Doi Numarası: 10.1016/j.nima.2003.09.066
- Dergi Adı: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
- Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
- Sayfa Sayıları: ss.454-467
- Anahtar Kelimeler: pulse height weighting technique, C6D6 scintillation detectors, Monte Carlo simulations, neutron capture cross-sections
- Bursa Uludağ Üniversitesi Adresli: Hayır
Özet
The accuracy of the pulse height weighting technique for the determination of neutron capture cross-sections is investigated. The technique is applied to measurements performed with C6D6 liquid scintillation detectors of two different types using capture samples of various dimensions. The data for well-known (n, gamma) resonances are analyzed using weighting functions obtained from Monte Carlo simulations of the experimental set-up. Several causes of systematic deviation are identified and their effect is quantified. In all the cases measured the reaction yield agrees with the standard value within 2%. (C) 2003 Elsevier B.V. All rights reserved.