Application of factorial design for mercury determination by trapping and graphite furnace atomic absorption spectrometry


Izgi B., Demir C., Gucer S.

SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, cilt.55, sa.7, ss.971-977, 2000 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 55 Sayı: 7
  • Basım Tarihi: 2000
  • Doi Numarası: 10.1016/s0584-8547(00)00234-2
  • Dergi Adı: SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.971-977
  • Anahtar Kelimeler: atomic absorption spectrometry, graphite furnace, blood serum, mercury determination, factorial design, OPTIMIZATION, AMALGAMATION, GOLD
  • Bursa Uludağ Üniversitesi Adresli: Evet

Özet

A trapping technique was developed for the determination of mercury traces in blood serum by graphite furnace atomic absorption spectrometry (GFAAS). The mercury was converted to metallic vapour in a reaction cell by tin chloride solution, swept by nitrogen into a gold-plated graphite tube and subsequently atomized. The influence of four experimental factors on the sensitivity of mercury determination was examined using two level factorial designs. It was found that the solution volume is the most significant parameter. (C) 2000 Elsevier Science B.V. All rights reserved.