Direct atomic absorption determination of mercury in drinking water and urine using a two-step electrothermal atomizer


Vil'pan Y., Grinshtein I., Akatov A., Gucer S.

JOURNAL OF ANALYTICAL CHEMISTRY, cilt.60, sa.1, ss.38-44, 2005 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 60 Sayı: 1
  • Basım Tarihi: 2005
  • Doi Numarası: 10.1007/s10809-005-0045-6
  • Dergi Adı: JOURNAL OF ANALYTICAL CHEMISTRY
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.38-44
  • Bursa Uludağ Üniversitesi Adresli: Hayır

Özet

A new method was developed for the direct electrothermal atomic absorption determination of mercury in drinking water and urine using double vaporization in a two-step atomizer with a purged vaporizer. In this method, a sample is placed in the vaporizer of a two-step atomizer, dried, and vaporized. The sample vapor is transferred to an unheated atomizer cell with a flow of argon and trapped by the inner surface of cell walls. This procedure can be performed repeatedly to preconcentrate mercury in the atomizer cell. Next, a portion of the sample transferred to the inner surface of the atomizer cell is revaporized and atomized by heating the atomizer cell of the two-step atomizer with a purged vaporizer, and the atomic absorption of mercury is measured. It was found that the degree of mercury transfer and trapping is as high as 100% at sufficiently high temperatures of primary vaporization, regardless of the material of the inner surface of the atomizer cell. The detection limits for mercury were 0.24 or 0.024 mug/L for drinking water at a sample volume of 100 muL using a single sample transfer or the procedure repeated ten times, respectively, and 2.0 mug/L for urine at a sample volume of 20 muL and a single sample transfer. The accuracy of the results was confirmed by the analysis of certified mercury samples and samples with known additives.